Can you set up any RIGOL scopes to stop triggering after a specific averaging number?

Last Updated: Dec 30, 2015 04:22PM PST
       Can you set up any RIGOL scopes to stop triggering after a specific averaging number?           

The average acquisition mode for the DS1000E/D, DS1000CA, DS1000B, DS4000, and DS6000 scopes are identical in function. They collect a boxcar average of the last n (usually a power of 2)  acquisitions on a rolling basis.

On the DS1000Z, DS2000/A, DS4000, and DS6000 scopes users can enable record mode to capture a specific amount of triggers into a set of frames, but to average all that data together it would have to be done externally.