What is the resolution of data returned by the DS1000E/D series of scopes?

Last Updated: Dec 30, 2015 05:09PM PST
All data returned from the DS1000E/D series have three significant digits.

It is all 8 bit data. 

This is true of all of our 8 bit scopes. Some of our scopes have a high resolution mode that can be enabled on the front panel, but this does not get returned to the PC with more significant digits. It appears as lower noise (via oversampling) on the front panel display only.