DS4000 Scope Calibration with the 5 to 1 BNC Cable

Last Updated: May 29, 2019 02:34PM PDT
Performing a Self Cal with the DS6000-CK 5 to 1 BNC Calibration Fixture
 
A Self Calibration can be performed on the 4000 series scopes.

To perform the self cal push the Utility key on the keypad and then choose Self-Cal in the list of available items. The self cal can then be started by pushing Start in the Self Cal menu.



Before a self calibration is performed you will need to connect a 5 to 1 BNC cable to each of the 5 BNC inputs on the front of the scope and the Trig Out/Calibration BNC at the back of the scope.

Here is a picture of the 5 to 1 BNC cable available from Rigol. It can be ordered by calling 877-474-4651 and asking for the DS6000-CK:

 

One can also be made by joining a group of five equal length (12" works well) to a single (20") BNC cable:



 